ONLINE MACHINE

  • 产品名称: Laser Radiation Monitoring System
  • NO.: NOVA II
  • Views : 2

 Confirm to IEC60950、IEC60825-2、IEC60825-12.

 

 Used to measure the laser wavelength, aperture, power generated by laser or LED.

 

 Laser or LED components.

 

 

OPHIR Laser power meter probe PD300R-3W

Detector Type

Silicon

Aperture

φ10mm

Filter Mode

Filter Out

Filter In
Spectral Range nm 350~1100 430~1100
Power Range 100mW to 5nW 100mW to 5nW
Power Scales 100mW to 300nW and dBm 100mW to 300nW dBm
Resolution nW 0.1 NA
Maximum Power vs.
Wavelength
nm mW mW
  <488 100 3000
  633 100 3000
  670 100 2000
  790 100 1200
  904 100 1200
  1064 100 2000
Accuracy (including errors due to temp. variations)
% error vs Wavelength nm ±10  360-40 NA
  ±3  400-950 ±5  430-950
  ±5  950-1100 ±7  950-1100
Damage Threshold W/cm² 10 100
Max Pulse Energy μJ 20 500
Noise Level for filter out pW 200  
Response Time with Meter s 0.2  
Beam Position Dependence ±2% ±3%
Fiber Adapters Available (see page 63) FC, ST, SC, SMA

 